Direct in-scope suction: an in vitro evaluation of a single use flexible ureteroscope with......
发布时间:2024-09-01 阅读次数:85

文献封面240901.png


Authors

Aideen Madden, Paris, France; Carlos Altez, Coruna, Spain; Jordi Pena Lueza, Lleida, Spain; Razvan Popescu, Bucharest, Romania; Johan Cabrera, Mariela Corralles, Olivier Traxer, Paris, France



Abstract


Purpose

To evaluate the stone clearance rate and limitations of a novel integrated suction capability within a single-use flexible ureteroscope according to stone particle size.

Methods

Varying sized stone particles were created using a stone phantom (Begostone Plus, Bego ©, Lincoln, RI, USA). Particle size ranged as follows: 63–125 µm, 125–250 µm, 250–500 µm, 500 µm- 1 mm and 1–2 mm. These were mixed with Normal saline (0.9%) to mimic in vivo conditions. The suction enabled single-use flexible ureteroscope (Pusen, Zhuhai, China) was used to aspirate stone fragments in three trials. Firstly, aspiration of 5 g of each dust range was attempted. Secondly, a direct comparison of the integrated suction to a manual syringe technique was applied to 1 g of each size range. Finally, aspiration was applied to 5 g of a heterogenous 1:1:1:1:1 mixture. Endoscopic clearance rate (g/min) and number of blockages were recorded. Each challenge was repeated three times.

Results

The integrated suction cleared 100% of dust < 250 µm. Endoscopic clearance rates were significantly faster than manual aspiration (3.01 g/min versus 0.41 g/min) for dust between 125 and 250 µm (p = 0.008). Complete endoscopic clearance by 180 s (without encountering test limiting blockages) was unsuccessful for particles > 250 µm. Clearance rates were greatly limited by stone particle size heterogeneity above 250 µm, to 0.09 g/min.

Conclusion

This technology works better in vitro than previously DISS evaluated methods when challenged by stone dust < 250 µm. However, this adaptation is significantly challenged in the presence of stone dust particles > 250 µm.

Keywords

DlSS; Direct-in-scope suction; Suction; Ureteroscopy


阅读原文 >>>



上一篇:First clinical evaluation of a flexible digital ureteroscope with direct in scope suctioning......
产品中心
  • 外科产品线
  • 内科产品线
  • 配件
新闻中心
  • 会议要闻
  • 企业动态
学术前沿
  • 一次性输尿管软镜
  • 文献分享
走进普生
  • 发展历程
  • 行销全球
  • 公司荣誉
  • 大爱普生
联系我们
  • 联系我们
  • 招贤纳才
  • 我在普生

扫一扫,关注我们

了解更多最新资讯

广东省珠海市高新区科技三路33号厂房1五楼 

400-688-2023  

pusen@pusenmedical.com

©2025 珠海普生医疗科技股份有限公司  All Rights Reserved  
 粤网药信备
字〔2025〕第00816号  1710212326908180.png粤公网安备44049102496975号   

反馈收集单